Perfect for keeping slices viable for longer.

Description
The 5100-Plus is perfect for those who need to keep slices viable for longer e.g. for electrophysiological field recordings. The user can calibrate the Z-axis deflection of the blade to 2 µm with the adjustable blade holder and “Opti-cal” Calibration device. Minimal Z-axis deflection reduces the “chatter” and minimizes damage to the slice surface, leading to greater viability of the cells on the slice surface.
Features
- Includes the Z-axis calibration unit (5100mz-Plus only)
- Vibration frequency from 50 to 80 Hz
- Adjustable amplitude from 0.5mm to 1.5mm
- Fine control of the blade advance (min. 0.1 mm/sec)
- Set start and finish position of blade travel
- Remote controls allowing for sterile use in culture hood
- Ice water bath easily removed for cleaning
- Optional cold light and magnifier for clear observation
- Optional Peltier cooled bath
Specifications
- Section thickness step size: 0.001 mm
- Bath table rise & fall speed: 1.0 mm/sec maximum
- Maximum (vertical) travel of bath table: 19 mm
- Cutting head advance speed:
- Minimum: -2.0 mm/sec (-1.00 during slicing)
- Maximum: +2.0 mm/sec (+1.00 during slicing)
- Cutting head retraction speed: 2.0 mm/sec
- Maximum travel of cutting head: 40 mm
- Blade oscillation frequency
- Minimum: 50 Hz
- Maximum: 80 Hz (amplitude dependent)
- Frequency step size: 5 Hz
- Blade oscillation amplitude
- Minimum: 0.5 mm (nominal)
- Maximum: 1.5 mm (nominal)
- Amplitude step size: 0.5 mm (nominal)
- Power requirements (Selectable): 115VAC 60Hz, 230VAC 50Hz
- Power rating: 100W
- Fuse rating
- 115V: T2A 250VAC
- 230V: T2A 250VAC
- Light source: Powered from 5100mz
- Dimensions: 350mm Width x 450mm Depth x 350mm Height (excluding magnifier/light source)
- Weight: 15kg (excluding magnifier)
Downloads
Currently not available
References